首页 | 本学科首页   官方微博 | 高级检索  
     


Optimization of spoiler gradients in flash MRI
Authors:Michael L. Wood   Michael Silver  Val M. Runge
Affiliation:

a Department of Radiation Oncology, New England Medical Center, Boston, Massachusetts 02111, USA

b Department of Radiology, New England Medical Center, Boston, Massachusetts 02111, USA

c Tufts University School of Medicine, Boston, Massachusetts 02111, USA

d Siemens Medical Systems, Iselin, New Jersey 08830, USA

Abstract:The FLASH technique for fast magnetic resonance (MR) imaging often employs strong magnetic field gradients, called spoiler gradients, to dephase the transverse magnetization after it has been measured. Otherwise, image artifacts can develop. The effectiveness of spoiler gradients at suppressing these artifacts was evaluated experimentally on two-dimensional MR images of a uniform phantom and patients. It was informative to compare the magnetization immediately before the RF excitation in each phase encoding step. Only spoiler gradients in the slice selection direction were effective. Spoiler gradients that decreased steadily from a large amplitude in the first phase encoding step to zero in the last minimized the transverse magnetization and suppressed the image artifact, without changing the image contrast.
Keywords:Spoiler gradients   Fast MRI   MRI techniques