淋洗液自动发生-离子色谱法测定重整催化剂中氯的含量
引用本文:赵雅郡,谢莉,周勇,冯移丽.淋洗液自动发生-离子色谱法测定重整催化剂中氯的含量[J].分析化学,2011,39(3):429-431.
作者姓名:赵雅郡  谢莉  周勇  冯移丽
作者单位:石油化工科学研究院;
摘    要:建立了重整催化剂中氯含量的淋洗液自动发生-离子色谱测定方法.以4 mL 7.5 mol/L NaOH溶液提取重整催化剂中Cl-,以淋洗液自动发生装置产生的KOH溶液为流动相,进行离子色谱法测定.结果表明,溶液中Cl-线性关系良好,相关系数为0.9999,测定结果的RSD小于0.76%,加标回收率为98.7%~102.8...

关 键 词:离子色谱  重整催化剂  

Determination of Chlorine in Reforming Catalyst by Ion Chromatography with Eluent Autogeneration
ZHAO Ya-Jun,XIE Li,ZHOU Yong,FENG Yi-Li.Determination of Chlorine in Reforming Catalyst by Ion Chromatography with Eluent Autogeneration[J].Chinese Journal of Analytical Chemistry,2011,39(3):429-431.
Authors:ZHAO Ya-Jun  XIE Li  ZHOU Yong  FENG Yi-Li
Institution:ZHAO Ya-Jun,XIE Li,ZHOU Yong,FENG Yi-Li(Research Institute of Petroleum Processing,Beijing 100083)
Abstract:A method for determination of chlorine in reforming catalyst by ion chromatography with eluent autogeneration was established.Chlorine in reforming catalyst was extracted by 4 mL sodium hydroxide solution of 7.5 mol/L,and determined by ion chromatography using potassium hydroxide solution generated by eluent generator as mobile phase.The linear relation of chlorine of solution was good,the correlation coefficient was 0.9999.The relative standard deviation was less than 0.8%,the recoveries were 98.7% to 102....
Keywords:Ion chromatography  Reforming catalyst  Chlorine  