首页 | 本学科首页   官方微博 | 高级检索  
     


Anodic stripping coulometry with collection for absolute trace analysis of lead, using flow-through electrochemical cells with porous working electrodes
Authors:E. Beinrohr   M. Németh   P. Tsch?pel  G. T?lg
Affiliation:(1) Laboratorium für Reinststoffanalytik, Max-Planck-Institut für Metallforschung, Bunsen-Kirchhoff-Strasse 13, W-4600 Dortmund 1, Federal Republic of Germany;(2) Department of Analytical Chemistry, Slovak Technical University, CS-812 37 Bratislava, Czechoslovakia;(3) Institut für Spektrochemie und angewandte Spektroskopie, Bunsen-Kirchhoff-Straße 11, W-4600 Dortmund, Federal Republic of Germany
Abstract:Summary Flow-through electrochemical cells with porous working electrodes made of crushed reticulated vitreous carbon and plated with mercury were used for absolute analysis of trace amounts of lead by anodic stripping coulometry with collection (ASCWC) in a flow system. The role of mercury coating, flow rate and pH were investigated. The coulombic content of the collection peak corresponded to the theoretical values calculated by Faraday's law in a concentration range from about 10–9 to 10–6 mol/l. The relative error and the relative standard deviation was +0.15% and 0.8%, respectively for 2×10–6 mol/l analyte concentration. The absolute detection limit (3 s) was 0.1 ng of Pb, the linear response range 7×104.One leave from: Department of Analytical Chemistry, Slovak Technical University, CS-812 37 Bratislava, Czechoslovakia
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号