Analysis of aqueous 2,4,6-trinitrotoluene (TNT) using a fluorescent displacement immunoassay
Authors:Goldman Ellen R  Anderson George P  Lebedev Nikolai  Lingerfelt Brian M  Winter Phan T  Patterson Charles H  Mauro J Matthew
Institution:US Naval Research Laboratory, Center for Bio/Molecular Science and Engineering, DC 20375, Washington, USA. erg@cbmse.nrl.navy.mil
Abstract:We report a rapid, simple, and sensitive assay that is potentially amenable to high throughput screening for analysis of 2,4,6-trinitrotoluene (TNT) present in aqueous solutions. The assay is based on the change in fluorescence emission intensity of a fluorescently labeled TNT analogue pre-bound to an anti-TNT antibody that occurs upon its competitive displacement by TNT. The assay can be performed in both cuvette- and 96-well plate-based formats. TNT at a level of 0.5 micro g L(-1) (0.5 ppb) was detected in phosphate buffered saline; detection improved to 0.05 micro g L(-1) (0.05 ppb) for TNT dissolved in artificial seawater.