首页 | 本学科首页   官方微博 | 高级检索  
     


A high performance system to study the influence of temperature in on-line solid-phase extraction capillary electrophoresis
Authors:Marcos Tascon  Fernando Benavente  Victoria Sanz-Nebot  Leonardo G. Gagliardi
Affiliation:1. Laboratorio de Investigación y Desarrollo de Métodos Analíticos (LIDMA), Universidad Nacional de La Plata, Calle 47 y 115 (1900), La Plata. Buenos Aires, Argentina;2. Centro de Investigación y Desarrollo en Tecnología de Pinturas (CIDEPINT-CIC-CONICET), Av. 52 y 120 (1900), La Plata, Buenos Aires, Argentina;3. Departamento de Química Analítica, Universidad de Barcelona, Av. Diagonal 645, Barcelona, España
Abstract:A novel high performance system to control the temperature of the microcartridge in on-line solid phase extraction capillary electrophoresis (SPE–CE) is introduced. The mini-device consists in a thermostatic bath that fits inside of the cassette of any commercial CE instrument, while its temperature is controlled from an external circuit of liquid connecting three different water baths. The circuits are controlled from a switchboard connected to an array of electrovalves that allow to rapidly alternate the water circulation through the mini-thermostatic-bath between temperatures from 5 to 90 °C. The combination of the mini-device and the forced-air thermostatization system of the commercial CE instrument allows to optimize independently the temperature of the sample loading, the clean-up, the analyte elution and the electrophoretic separation steps.
Keywords:Capillary electrophoresis   On-line pre-concentration   Opioid peptides   Solid-Phase extraction   Temperature
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号