首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Spectrochemical determination of trace As, Fe, Hg, Mn, and Pb by an argon-stabilized U-shaped DC arc
Authors:D M Markovi?  I Novovic?
Institution:(1) Institute of Physics, P.O. Box 68, 11080 Belgrade, Pregrevica 118, Serbia
Abstract:Argon-stabilized U-shaped DC arc with a system for aerosol introduction was used for the determination of As, Fe, Hg, Mn, and Pb. By applying a computer program performing a time integration of the registered signals and by optimizing the working conditions of an U-shaped arc plasma, the detection capability for As, Fe, Hg, Mn, and Pb was improved, which contributed to the lowering of the detection limits. The lowest detection limits for Fe, Mn, and Pb were achieved during an integration time of 20 s, and for As and Hg during an integration time of 30 s, and their values were 1.1, 0.1, 0.9, 15, and 2.6 ng/mL, respectively. The detection limits obtained by our method with optimal integration times were compared with the detection limits obtained using other methods: inductively coupled plasma-atomic emission spectrometry (ICP-AES), directly coupled plasma-atomic emission spectrometry (DCP-AES), microwave-induced plasma-atomic emission spectrometry (MIP-AES), inductively coupled plasma-mass spectrometry (ICP-MS), and an improved thermospray flame furnace atomic absorption spectrometry (TS-FF-AAS). The text was submitted by the authors in English.
Keywords:
本文献已被 SpringerLink 等数据库收录!