Integration of a contactless conductivity detector into a commercial capillary cassette. Detection of inorganic cations and catecholamines |
| |
Authors: | Vuorinen Pasi S Jussila Matti Sirén Heli Palonen Sami Riekkola Marja-Liisa |
| |
Affiliation: | Laboratory of Analytical Chemistry, Department of Chemistry, University of Helsinki, P.O. Box 55, FIN-00014, University of Helsinki, Finland |
| |
Abstract: | A contactless conductivity detector integrated into the capillary cassette of Agilent 3DCE equipment is described. The detector is user-friendly, compact and easily modified. The UV detector of the 3DCE equipment is available parallel with the contactless conductivity detector increasing the detection power. Two electrolyte solutions, 2-(N-morpholino)ethanesulfonic acid–histidine solution (20 mM, pH 6.0) and ammonium acetate (10 mM, pH 4.0), were used as the separation media for inorganic cations and organic catecholamines, respectively. The detection limit for all metal cations except barium was under 0.5 mg/l, and that for four catecholamines was ca. 10 mg/l. This last value was the same order of magnitude as achieved with parallel UV detection. |
| |
Keywords: | Inorganic cations Catecholamines |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|