High-throughput protein analysis by multiplexed sodium dodecyl sulfate capillary gel electrophoresis with UV absorption detection |
| |
Authors: | Luo Siquan Feng Jianmin Pang Ho-Ming |
| |
Affiliation: | CombiSep, Inc., 2711 South Loop Drive, Suite 4200, Ames, IA 50010, USA. |
| |
Abstract: | We have developed a novel, high-throughput approach for molecular mass determination of proteins from 14 200 to 116 000 based upon multiplexed, absorbance-based capillary electrophoresis. Via capillary multiplexing, 96 samples were analyzed simultaneously within 30 min. Detection with ultraviolet light obviates the need for protein staining or derivatization. The detection limit of the system was estimated at 5 microg/ml bovine serum albumin (BSA) when sampled from 12.5 mM Tris-HCl. The linear dynamic range was over two orders of magnitude from 5 microg/ml to 1000 microg/ml for BSA. Better than 5% sizing accuracy for protein molecular mass determination and excellent run-to-run and day-to-day reproducibility was obtainable with the described method. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|