Determination of total iodine in foods and dietary supplements using inductively coupled plasma-mass spectrometry
Authors:Sullivan Darryl  Zywicki Richard
Institution:Covance Laboratories Inc., 3301 Kinsman Blvd, Madison, WI 53704, USA. darryl.sullivan@covance.com
Abstract:A method was developed and validated for the determination of total iodine in a wide variety of food products and dietary supplements. The method involves a unique sample digestion with a KOH solution in an oven or by using an open-vessel microwave system. After digestion, a stabilizer is added and the solution is taken to volume, then filtered and analyzed either directly or after dilution. The amount of iodine is determined with inductively coupled plasma-mass spectrometry (ICP-MS). The method was validated by experiments to determine its precision, accuracy, linearity, specificity, ruggedness, and robustness. The LOQ of this method is 25-50 microg/kg. The method demonstrated an average RSD of 2.27% during analysis of milk powder and 4.30% during analysis of a dietary supplement tablet reference material. The accuracy of the method as determined with these same reference materials was 100 and 94.2%, respectively. The method has been used successfully on commodity foods, processed foods, dairy products, pet food, infant formula, animal feed, mineral premixes, and a variety of dietary supplements.
Keywords:
本文献已被 PubMed 等数据库收录!