首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Donnan dialysis preconcentration coupled with ion chromatography and electrocatalytic oxidation for the determination of cyanide
Authors:Seneviratne J  Holmstrom S D  Cox J A
Institution:

Department of Chemistry and Biochemistry, Miami University, Oxford, OH 45056, USA

Abstract:An electrocatalytic amperometric detector for the ion chromatographic determination of CN? is described. A conducting composite that is based on a graphite-loaded sol–gel material comprises the working electrode. The composite is doped with a RuII metallodendrimer which is demonstrated to promote the electrochemical oxidation of CN? at potentials positive of 0.5 V vs. Ag/AgCl. In 6 mM NaOH, 0.05 M NaCl flowed at 1.0 ml min?1, a 5-point calibration curve with the following linear least squares parameters is obtained over the range, 1.0–30 M CN?: slope, 24.2±0.1 nA M?1; intercept, ?6±2 nA; and r, 0.9997. The detection limit, 0.7 μM CN?, compares favorably to that obtained by amperometry at a silver electrode, 0.5 μM CN?, under comparable experimental conditions. A 60-min preconcentration by Donnan dialysis increases the sensitivity by a factor of 23.6.
Keywords:Cyanide  Ion chromatography  Amperometry  Electrocatalysis
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号