Abstract: | The development of methods for temperature and heat calibration of DSC and DTA supported by the German Society of Thermal Analysis (GEFTA) are presented. Reasons are given for the recommendation of the temperature calibration method, which refers to the valid international temperature scale (ITS 90) and takes into account the fact that the fixpoints of this scale are defined in two phase equilibrium.Furthermore the hard problems with the heat flow and peak area calibration of DSC are described and the methods are presented to come to realistic and sufficient results. Thus both traditional (with the aid ofCp — and enthalpy — reference materials) and electrical calibration are critically discussed. Today's best values of suitable materials for heat calibration are reported.
Zusammenfassung Methoden zur Temperatur- und Enthalpiekalibrierung von DSC und DTA Instrumenten werden dargestellt. Dieses Forschungsvorhaben wurde von der Gesellschaft für Thermische Analyse (GEFTA) unterstützt. Gründe werden dargelegt, weshalb die Temperaturkalibrierung auf Referenzsubstanzen abgestützt werden soll, die in der Internationalen Temperaturskala (ITS90) aufgeführt sind.Ebenso wird das schwierige Unterfangen der kalorischen Kalibrierung mit mehreren unterschiedlichen Methoden erlÄutert, und geeignete Referenzsubstanzen werden vorgestellt. |
| |