Use of mobile phase 18-crown-6 to improve peak resolution between mono- and divalent metal and amine cations in ion chromatography |
| |
Authors: | Richens Dean A Simpson David Peterson Sarah McGinn Arlo Lamb John D |
| |
Affiliation: | Department of Chemistry and Biochemistry, Brigham Young University, C100 Benson Science Building, P.O. Box 25700, Provo, UT 84602, USA. |
| |
Abstract: | It is difficult to quantify NH4+ by ion chromatography in the presence of high concentrations of Na+ due to peak overlap. The Dionex IonPac CS15 column, which contains phosphonate, carboxylate, and 18-crown-6 functional groups, was originally developed to overcome this problem. We have found that the addition of 18-crown-6 to the eluent promotes improved peak resolution between Na+ and NH4+ even at concentrations as high as 60,000 to 1 using this column. Its use also improves the separation of alkali and alkaline earth metal and amine cations. Mobile phase 18-crown-6 increased the retention times of CH3NH3+, NH4+, and K+, and decreased the retention time of Sr2+. The retention times of Li+, Na+, Mg2+, Ca2+, (CH3)2NH2+, and (CH3)3NH+ were not affected. This method makes possible the direct analysis of ammonia from nitrogenase, the enzyme responsible for biological nitrogen fixation. The resolution of the NH4+ peak from the Na+ and Mg2+ peaks improved from zero resolution to values of 6.19 and 5.65, respectively. This technique considerably reduces the analysis time of NH4+ in the presence of high concentrations of Mg2+ and Na+ over traditional indophenol measurements. |
| |
Keywords: | Peak resolution Mobile phase composition Ion chromatography 18-Crown-6 Sodium Ammonium Nitrogenase |