首页 | 本学科首页   官方微博 | 高级检索  
     


Impedimetric Aptasensor Based on Disposable Graphite Electrodes Developed for Thrombin Detection
Authors:Ece Eksin  Arzum Erdem  Ayben Pınar Kuruc  Hilal Kayi  Aylin Öğünç
Affiliation:1. Analytical Chemistry Department, Faculty of Pharmacy, Ege University, 35100, Bornova, Izmir, Turkey, Phone: +90‐232‐311?5131, Fax: +90‐232‐388?5258;2. Izmir Private Turkish College, Science High School, Mithatpasa Street, No.?689, Kopru‐Izmir, Turkey
Abstract:The impedimetric aptasensor for Thrombin (THR) was developed for the first time herein by measuring changes at the charge‐transfer resistance, Rct upon to protein? aptamer complex formation. After covalent activation of pencil graphite electrode (PGE) surface using covalent agents, amino linked aptamer (APT) was immobilized onto activated PGE surface. Then APT‐THR interaction was explored by electrochemical impedance spectroscopy (EIS). After the optimization of experimental conditions (e.g., APT and THR concentration, immobilization and interaction times), the selectivity of impedimetric aptasensor was tested in the presence of other biomolecules: factor Va and bovine serum albumine (BSA) both in buffer media, or in diluted fetal bovine serum (FBS).
Keywords:Thrombin  Aptamer  Electrochemical impedance spectroscopy  Pencil graphite electrode  Fetal bovine serum
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号