首页 | 本学科首页   官方微博 | 高级检索  
     


Microfluidic devices with disposable enzyme electrode for electrochemical monitoring of glucose concentrations
Authors:Li Xin  Zhang Fan  Shi Jian  Wang Li  Tian Jing-Hua  Zhou Xiong-Tu  Jiang Lian-Mei  Liu Li  Zhao Zhen-Jie  He Pin-Gang  Chen Yong
Affiliation:Ecole Normale Supérieure, Paris, France.
Abstract:This article describes the fabrication of tube-like microchannels made of UV curable polymer on a glass substrate and the device assembling with a disposable enzyme-working electrode for high-sensitivity electrochemical detection. While both reference and counter electrodes are patterned on the surface of the glass substrate, the working electrode is flipped on the top of the channel with an open access, providing a face-to-face probing configuration. When the enzyme electrode is contaminated or degraded, it can be easily replaced by a new one, keeping the main body of the device and the detection schema unchanged. Using glucose oxidase-coated gold electrodes, we were able to determine a linear amperometry response to the glucose concentrations in the range of 2-16 mM. By replacing the as-prepared working electrode by the one after thermal treatments, we showed a much more degraded enzyme electrode activity, enabling efficient determination of the electrode quality as well as the whole process optimization.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号