Affiliation:1. Fuqing Branch of Fujian Normal University, Fuqing, Fujian 350300, China;2. Ministry of Education Key Laboratory of Analysis and Detection for Food Safety, Fujian Provincial Key Laboratory of Analysis and Detection for Food Safety, and Department of Chemistry, Fuzhou University, Fuzhou, Fujian 350002, China
Abstract:An electrically heated microelectrode chip (HMEC) was designed and the Ru(bpy)32+/carbofuran electrochemiluminescence (ECL) systems were applied to characterize the performance of the HMEC. The ECL intensities increase at elevated electrode temperature, and the detection limit at 60°C (electrode surface temperature) was about 10 times lower than that at 30°C. The results indicate that new heated electrode can be handled easily and can be mass produced, the difference between the electrodes was little. The stability of the HMEC was good since the electrode surface can hardly be destroyed during detection and storage.
Keywords:electrically heated microelectrode chip  electrochemistry  luminescence  electrochemiluminescence  carbofuran