Flow-injection Determination of Cysteine,N-Acetyl Cysteine and Glutathione in Pharmaceuticals via Potassium Ferricyanide-Fe(Ⅲ) Spectrophotometric System |
| |
摘 要: | A simple flow injection spectrophotometric method is reported for the determination of cysteine,N-acetyl cysteine and glutathione based on the reduction of Fe(Ⅲ)/ferricyanide,the in situ reduced ions are reacted with unreduced portion of ferricyanide/Fe(Ⅲ) to form soluble Prussian blue,which is monitored at 735 nm.The calibration graphs are linear in the concentration ranges of(1―100)×10-6 mol/L for cysteine and N-acetyl cysteine,and(1―50)×10-6 mol/L for glutathione.The relative standard deviations of 1.8%,2.5% and 1.9% were found for eleven replicate analyses of 5×10-6 mol/L cysteine,N-acetyl cysteine and glutathione.The limits of detection(3σ blank) at 5×10-7 mol/L for cysteine,and 3×10-7 mol/L for N-acetyl cysteine and glutathione were obtained.The proposed method allowed 60 injections/h.The effects of common substances present in pharmaceuticals and human physiological fluids were examined.The method was applied to determining cysteine in pharmaceutical formulations with the recoveries in a range of 97% to 106% and the results obtained are agreed well with labeled values.
|
本文献已被 CNKI 等数据库收录! |
|