首页 | 本学科首页   官方微博 | 高级检索  
     


A type of spectrophotometer with both SCI and SCE measurement structures
Authors:Yuan Kun  Yan Huimin  Jin Shangzhong
Affiliation:1. State Key Laboratory of Modern Optical Instrument, Zhejiang University, Hangzhou, Zhejiang 310027, China;2. College of Optical and Electronic Technology, China Jiliang University, Hangzhou, Zhejiang 310018, China
Abstract:In order to measure the color of different samples with different surface texture, multiple measurement and observation conditions are needed from the instrument. Thus, both SCI (specular component included) and SCE (specular component excluded) measurement conditions are included in most color measuring instruments today. In current design, a light trap is established in the integrating sphere, and the instrument uses mechanical control to switch between SCI and SCE measurement modes. However, establishing a light trap inside the integrating sphere will affect the SCI and SCE measurement results. This article discusses the problems in current design, proposes a dual light source structure that includes both SCI and SCE measurement conditions, and then evaluates the structure with experiments. The experiment results show that the improved structure is more suitable for color measuring instruments with both SCI and SCE modes, and it has high practical value.
Keywords:120.2440   120.4640   120.5240
本文献已被 ScienceDirect 等数据库收录!