首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Simultaneous determination of cotinine and trans‐3‐hydroxycotinine in urine by automated solid‐phase extraction using gas chromatography–mass spectrometry
Authors:Fouad Chiadmi  Joël Schlatter
Institution:Laboratoire de toxicologie de médecine légale, H?pital Jean Verdier – APHP, Bondy, France
Abstract:A gas chromatography–mass spectrometry method was developed and validated for the simultaneous automated solid‐phase extraction and quantification of cotinine and trans‐3‐hydroxycotinine in human urine. Good linearity was observed over the concentration ranges studied (R2 > 0.99). The limit of quantification was 10 ng/mL for both analytes. The limits of detection were 0.06 ng/mL for cotinine (COT) and 0.02 ng/mL for trans‐3‐hydroxycotinine (OH‐COT). Accuracy for COT ranged from 0.98 to 5.28% and the precision ranged from 1.24 to 8.78%. Accuracy for OH‐COT ranged from ?2.66 to 3.72% and the precision ranged from 3.15 to 7.07%. Mean recoveries for cotinine and trans‐3‐hydroxycotinine ranged from 77.7 to 89.1%, and from 75.4 to 90.2%, respectively. This analytical method for the simultaneous measurement of cotinine and trans‐3‐hydroxycotinine in urine will be used to monitor tobacco smoking in pregnant women and will permit the usefulness of trans‐3‐hydroxycotinine as a specific biomarker of tobacco exposure to be determined. © 2014 The Authors. Biomedical Chromatography published by John Wiley & Sons Ltd.
Keywords:cotinine  trans‐3‐hydroxycotinine  urine  GCMS  SPE
设为首页 | 免责声明 | 关于勤云 | 加入收藏