Determination of concentrated hydrogen peroxide at single-walled carbon nanohorn paste electrode
Institution:1. State Key Laboratory of Electroanalytical Chemistry, Changchun Institute of Applied Chemistry, Chinese Academy of Sciences, Changchun Jilin 130022, China;2. Graduate University of the Chinese Academy of Sciences, Chinese Academy of Sciences, Beijing 100864, China
Abstract:Single-walled carbon nanohorn (SWCNH) paste electrode was used for amperometric determination of concentrated hydrogen peroxide, and was compared with other carbon electrodes. The calibration plots are linear from 0.5 to 100 mM at activated SWCNH paste electrode and edge plane graphite (EPG) electrode. In contrast, the calibration plots are linear only at concentrations lower than 45 mM at graphite paste electrode, multi-walled carbon nanotube paste electrode, and glassy carbon electrode. Our results show that SWCNH paste electrode and EPG electrode are interesting alternatives to high surface area platinum electrode for determination of concentrated hydrogen peroxide. Because of its high-purity, metal-free SWCNH is a user-friendly and attractive material for electrochemical study.
Keywords:
本文献已被 ScienceDirect 等数据库收录!