首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of arsenic in ores, concentrates and related materials by continuous hydride-generation atomic-absorption spectrometry after separation by xanthate extraction
Authors:Donaldson E M  Leaver M E
Affiliation:Mineral Sciences Laboratories, Canada Centre for Mineral and Energy Technology, Department of Energy, Mines and Resources, Ottawa, Canada.
Abstract:A recent graphite-furnace atomic-absorption method for determining approximately 0.2 mug/g or more of arsenic in ores, concentrates, rocks, soils and sediments, after separation from matrix elements by cyclohexane extraction of arsenic(III) xanthate from approximately 8-10M hydrochloric acid, has been modified to include an alternative hydride-generation atomic-absorption finish. After the extract has been washed with 10M hydrochloric acid-2% thiourea solution to remove co-extracted copper and residual iron, arsenic(III) in the extract is oxidized to arsenic(V) with bromine solution in carbon tetrachloride and stripped into water. Following the removal of bromine by evaporation of the solution, arsenic is reduced to arsenic(III) with potassium iodide in approximately 4M hydrochloric acid and ultimately determined to hydride-generation atomic-absorption spectrometry at 193.7 nm, with sodium borohydride as reductant. Interference from gold, platinum and palladium, which are partly co-extracted as xanthates under the proposed conditions, is eliminated by complexing them with thiosemicarbazide before the iodide reduction step. The detection limits for ores and related materials is approximately 0.1 mug of arsenic per g. Results obtained by this method are compared with those obtained previously by the graphite-furnace method.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号