Coulometric titration with gas diffusion separation step |
| |
Authors: | M. Hahn H. H. Rüttinger H. Matschiner |
| |
Affiliation: | (1) Institute for Analytical Chemistry, Martin Luther University Halle, Weinbergweg 16, O-4020 Halle/S., Federal Republic of Germany |
| |
Abstract: | Summary Coulometric titration is employed in combination with a gas diffusion step for the quantitative separation and determination of volatile substances. Complete separation is brought about by filling the sample into a closed circulating loop and transporting it repeatedly past a microporous membrane. The combination of membrane separation module with the coulometric cell allows rapid separation and exploitation of the precise absolute determinations obtainable by the coulometric titration method. The high precision and selectivity of the determination and quantitative separation are discussed using ammonia and sulphur dioxide analyses as examples.Part I: Fresenius Z Anal Chem (1989) 332:849–854 |
| |
Keywords: | |