首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Analysis of volatile organic chemicals in aqueous samples by purge/GC with selective water removal
Authors:J W Cochran  J M Henson
Abstract:A gas chromatographic method for volatile organic chemicals in which an aqueous sample is purged directly to a cryogenically cooled, fused silica column uses a Nafion tube drier between the purge vessel and GC column. The Nafion strips water from the gas stream during the purge step while allowing volatile halocarbons and aromatics to continue to the GC column. Examples of this technique are presented on 0.53 mm and 0.25 mm fused silica columns coated with a variety of stationary phases.
Keywords:Capillary chromatography  Purge GC  Volatile organic chemicals
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号