氢化物发生-原子荧光光谱法对水中总硒含量测试方法改进研究
作者姓名:Xu YX  Luo KL
作者单位:中国科学院地理科学与资源研究所;中国科学院研究生院
基金项目:国家自然科学基金项目(40872210);国家(863计划)高技术研究发展计划项目(2004AA601080;2006AA06Z380);陕西省紫阳县科技计划项目资助
摘    要:为探寻简单可靠的水中硒测试方法,在采用国标中氢化物发生-原子荧光光谱法测定水样中的硒含量时,对影响水中硒消解的因素如消解温度和盐酸用量等进行了比较,并且确定了一种简易可行的新方法,即水样在100℃消解时加入3mL盐酸。采用3,6和10μg.L-1硒标液测定了方法的精密度和准确度,测试的结果分别是2.96,5.43和9.66μg.L-1,回收率在90.50%~98.67%。同时,用不同方法测试了29个水样中的硒含量,国标中氢化物发生-原子荧光光谱法测定水样中的硒含量的标样的误差为8.95%~25.46%,新方法测定标样的误差为1.33%~3.40%。新方法相对于中华人民共和国标准GB/T 5750.6-2006水中硒测试方法而言,具有操作过程简单、易控制和可靠性强、经济等特点。

关 键 词:荧光光谱法  测试方法  水中总硒  改进

Determination of total selenium in water by hydride generation atomic fluorescence spectrometry
Xu YX,Luo KL.Determination of total selenium in water by hydride generation atomic fluorescence spectrometry[J].Spectroscopy and Spectral Analysis,2012,32(2):532-536.
Authors:Xu Yong-xin  Luo Kun-li
Institution:Institute of Geographic Sciences and Natural Resources Research, Chinese Academy of Sciences, Beijing 100049, China. xyxn126@126.com
Abstract:In order to establish a rapid and accurate hydride generation-atomic fluorescence spectrometry method for the determination of selenium content in water samples, the influence factors on Se nitrolysis in the water such as nitrolysis temperature and the amount of hydrochloric acid were studied. The authors determined a simple and practical test method that is to add 3 mL hydrochloric acid at 3 degrees C. We measured the precision and accuracy of the method by the standard liquid of 3, 6 and 10 micro x L(-1), and the test results were 2.96, 5.43 and 9.66 mg x L(-1). The recovery varied from 90.50% to 98.67%. Meanwhile we tested 29 water samples by different methods. The error of standard sample was 8.95%-25.46% by hydride generation-atomic fluoresence spectrometry in the GB/T 5750. 6-2006, and the error of standard sample was 1.33%-3.40% by new method. Compared with the national standards of GB/T 5750.6-2006, the method has the advantages of operating simplicity and and easy control, and it is also reliable and economical.
Keywords:Fluorescence spectroscopy  Test method  Toal selenium in water  Improved
本文献已被 CNKI PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号