首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Effects of angle of probe laser and pinhole diameter on the time-resolved optical inspection system
Authors:Chil-Chyuan Kuo
Institution:Department of Mechanical Engineering, Ming Chi University of Technology, No. 84, Gungjuan Road, Taishan Taipei Hsien 243, Taiwan
Abstract:A real-time in-situ time-resolved optical inspection system comprising two He-Ne probe lasers, a digital oscilloscope, and three fast photodetectors is developed. The effects of angle of probe laser and pinhole diameter on the time-resolved optical inspection system are investigated during excimer laser crystallization. The longest melt duration and ablation excimer laser fluence are independent of pinhole size. The pinhole with a diameter of 0.3 mm is recommended to use in the time-resolved optical inspection system. Determination of melt duration and observation of explosive crystallization oscillation are independent of angle of probe laser, but four criteria for the relationship between probe lasers and photodetectors should be followed.
Keywords:Optical inspection system  Angle of probe laser  Pinhole diameter
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏