In situ measurement of transuranics using a calcium fluoride scintillation detection system |
| |
Authors: | M J Rudin W M Richardson P G Dumont W H Johnson |
| |
Institution: | (1) Department of Health Physics, University of Nevada, Las Vegas, 4505 Maryland Parkway, Las Vegas, NV 89154, USA;(2) Aptec-NRC, 125 Titus Avenue, Warrington, PA 18976, USA |
| |
Abstract: | A prototype scintillation spectrometer capable of measuring low energy photons from transuranic elements in groundwater has been developed for in situ applications. The detector is a CaF2 (Eu) crystal enclosed in a stainless steel casing with a 19 mg·cm-2 stainless steel entrance window. Evaluation of the system was performed in the laboratory by submersing the probe into a number of aqueous solutions containing varying concentrations of 241Am in a simulated well environment. Spectra obtained had easily discernible peaks associated with the low energy L X-rays and the 60 keV gamma-ray photopeak of 241Am. |
| |