Potentiometric stripping determination of heavy metals using a graphite-reinforcement carbon vibrating electrode
Authors:T Kakizaki and Kiyoshi Hasebe
Institution:(1) Department of Chemistry, Hokkaido University of Education at Iwamizawa, Iwamizawa 068, Japan, JP;(2) Graduate School of Environmental Earth Science, Hokkaido University, Sapporo 060, Japan, JP
Abstract:Potentiometric stripping determination of Cd, Cu and Zn using a vibrating electrode (VE) is presented. A simple VE was constructed by using a piezoelectric bimorph oscillator and an inexpensive graphite-reinforcement carbon (GRC) rod (a mechanical pencil). Experimental results obtained with the simple VE follow the equation valid for quantitative application of potentiometric stripping analysis (PSA) in large solutions. It was found that the GRC vibrating electrode is suitable for multielement trace analysis of small samples of 0.05 mL. The relative standard deviations for divalent metals are better than 2%. Received: 3 February 1997 / Revised: 2 June 1997 / Accepted: 7 June 1997
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏