首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Simplified column-switching technology for the determination of traces of anions in the presence of high concentrations of other anions
Authors:Huan Y  Mou S F  Liu K N  Rivielo J M
Institution:Research Center for Eco-Environmental Sciences. Academia Sinica, Beijing, China.
Abstract:A binary eluent model was proposed to study the retention behavior of trace anions in the presence of high concentrations of co-eluent anions. According to this model, the effect of the matrix on the retention of trace anions can be suppressed by using a high-concentration eluent. Based on the model, a simplified 'heart-cut' column-switching technology was proposed, in which the column-switching time window can be determined directly by standard solution of the analyte. By using this technology, a detection limit of 50 microg/l was obtained for nitrate in the presence of 5,000 mg/l chloride and 250 mg/l sulfate. The effect of the matrix on the concentration efficiency of weak acid anions was studied according to the result of phosphate. The possibility of the determination of weak acid anion by the 'heart-cut' technology was also discussed. The technology has been applied for the determination of nitrate in high salinity sea water, the spike recovery is in the range of 89.2-101.3%.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号