首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Multisyringe flow injection spectrophotometric determination of uranium in water samples
Authors:Jorge L Guzmán Mar  Leticia López Martínez  Pedro L López de Alba  Nancy Ornelas Soto  Víctor Cerdà Martín
Institution:1.Instituto de Investigaciones Científicas,Universidad de Guanajuato,Guanajuato,Mexico;2.Departamento de Química,Universidad de las Islas Baleares,Palma de Mallorca,Spain
Abstract:A multisyringe flow injection analysis method for the determination of uranium in water samples was developed. The methodology was based on the complexation reaction of uranium with arsenazo (III) at pH 2.0. Uranium concentrations were spectrophotometrically detected at 649 nm using a light emitting diode. Under the optimized conditions, a linear dynamic range from 0.1 to 4.0 μg mL−1, a 3σ detection limit of 0.04 μg mL−1, and a 10σ quantification limit of 0.10 μg mL−1 were obtained. The reproducibility (%) at 0.5, 2.5, and 4.0 μg mL−1 was 2.5, 0.9, and 0.6%, respectively (n = 10). The interference effect of some ions was tested. The proposed method was successfully applied to the determination of uranium in water samples.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号