首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Reveal Listeria 2.0 test for detection of Listeria spp. in foods and environmental samples
Authors:Alles Susan  Curry Stephanie  Almy David  Jagadeesan Balamurugan  Rice Jennifer  Mozola Mark
Institution:Neogen Corp., 620 Lesher Pl, Lansing, MI 48912, USA.
Abstract:A Performance Tested Method validation study was conducted for a new lateral flow immunoassay (Reveal Listeria 2.0) for detection of Listeria spp. in foods and environmental samples. Results of inclusivity testing showed that the test detects all species of Listeria, with the exception of L. grayi. In exclusivity testing conducted under nonselective growth conditions, all non-listeriae tested produced negative Reveal assay results, except for three strains of Lactobacillus spp. However, these lactobacilli are inhibited by the selective Listeria Enrichment Single Step broth enrichment medium used with the Reveal method. Six foods were tested in parallel by the Reveal method and the U.S. Food and Drug Administration/Bacteriological Analytical Manual (FDA/BAM) reference culture procedure. Considering data from both internal and independent laboratory trials, overall sensitivity of the Reveal method relative to that of the FDA/BAM procedure was 101%. Four foods were tested in parallel by the Reveal method and the U.S. Department of Agriculture-Food Safety and Inspection Service (USDA-FSIS) reference culture procedure. Overall sensitivity of the Reveal method relative to that of the USDA-FSIS procedure was 98.2%. There were no statistically significant differences in the number of positives obtained by the Reveal and reference culture procedures in any food trials. In testing of swab or sponge samples from four types of environmental surfaces, sensitivity of Reveal relative to that of the USDA-FSIS reference culture procedure was 127%. For two surface types, differences in the number of positives obtained by the Reveal and reference methods were statistically significant, with more positives by the Reveal method in both cases. Specificity of the Reveal assay was 100%, as there were no unconfirmed positive results obtained in any phase of the testing. Results of ruggedness experiments showed that the Reveal assay is tolerant of modest deviations in test sample volume and device incubation time.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号