首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A chemiluminescence-based continuous flow aqueous ozone analyzer using photoactivated chromotropic acid
Authors:Takayanagi Toshio  Dasgupta Purnendu K
Institution:Department of Chemistry and Biochemistry, Texas Tech University, Lubbock, TX 79409-1061, USA
Abstract:Ozone has become the oxidant of choice for water disinfection, especially in large water treatment facilities. This paper describes a fast and sensitive method for the determination of ozone content by reaction with photoactivated chromotropic acid (CA, 4,5-dihydroxynaphthalene-2,7-disulfonic acid), which results in intense chemiluminescence (CL). Freshly ozonated water from a recirculating ozonizer/reservoir is injected into a carrier stream of deionized water in the flow-injection mode. This flow mixes with a stream of photoactivated CA solution in a spiral cell placed directly on top of an inexpensive miniature (8 mm diameter active area) photomultiplier tube (PMT). Alkaline CA is photoactivated by passing it through a FEP-Teflon® coil (residence time ∼ 50 s) wrapped around a 1 W UV lamp emitting at 254 nm; without photoactivation, the signal is ∼70-fold lower. The S/N = 3 limit of detection for aqueous ozone is 3 μg l−1 and good response slope is obtained up to an ozone concentration of 1.4 mg l−1, the highest that could be made in this study. The response obeyed a quadratic equation with r2 = 0.9984. No interference from permanganate ion is observed. The proposed system was applied to the monitoring of ozonation status of a playa lake water that exhibited significant ozone demand.
Keywords:Ozone  Photoactivation  Analyzer
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号