Fast on-line ultrasound-assisted extraction coupled to a flow injection-atomic absorption spectrometric system for zinc determination in meat samples |
| |
Authors: | Yebra-Biurrun M C Moreno-Cid A Cancela-Pérez S |
| |
Affiliation: | Department of Analytical Chemistry, Nutrition and Bromatology, Faculty of Chemistry, University of Santiago de Compostela, 15782 Santiago de Compostela, Spain |
| |
Abstract: | A continuous ultrasound-assisted extraction system connected to a flow injection manifold has been used for the on-line determination of zinc in meat samples by flame atomic absorption spectrometry. An experimental design was used for the optimisation of the continuous manifold. This flow injection methodology allowed a sampling frequency of ca. 80 samples per hour with a relative standard deviation for the whole procedure of 0.3% (for a sample containing 163.6 μg g−1 Zn). The detection limit was 0.6 μg g−1 for a sample amount of 5 mg. Accurate results were obtained by measuring certified reference materials (BCR-186 (pig kidney) and BCR-184 (bovine muscle)). The analytical procedure was applied to different real meat samples with satisfactory results. |
| |
Keywords: | Continuous ultrasound-assisted extraction system Flame atomic absorption spectrometry Zinc determination Meat samples |
本文献已被 ScienceDirect PubMed 等数据库收录! |