Comparison and joint use of near infrared spectroscopy and Fourier transform mid infrared spectroscopy for the determination of wine parameters |
| |
Authors: | Urbano Cuadrado M Luque de Castro M D Pérez Juan P M Gómez-Nieto M A |
| |
Affiliation: | a Department of Analytical Chemistry, University of Córdoba, Campus of Rabanales, Annex C3, E-14071 Córdoba, Spain b LIEC, Agrifood and R&D Laboratory, E-13200 Manzanares, Spain c Department of Computing and Numerical Analysis, University of Córdoba, Campus of Rabanales, Building C2, Floor 3, E-14071 Córdoba, Spain |
| |
Abstract: | A study of the statistic characteristics of the multidetermination of several enological parameters - namely, alcoholic degree, volumic mass, total acidity, glycerol, total polyphenol index, lactic acid and total sulphur dioxide - depending on the spectroscopic zone employed, was carried out. The two techniques used were near infrared spectroscopy (NIRS) and Fourier transform mid infrared spectroscopy (FT-MIRS). The combination of these two regions (sum of their spectra) was also studied. NIRS yielded better results, but the use of both zones improved the determination of glycerol and total sulphur dioxide. The training and validation sets used for developing general equations were built with samples from different apellation d’origine, different wine types, etc. Partial least squares regression was used for multivariate calibration, using systematic cross validation in the calibration stage and external validation in the testing stage. Sample preparation was not required. |
| |
Keywords: | Near infrared spectroscopy Attenuated total reflection mid infrared Wineries Multivariate calibration |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|