Elemental analyses of soil and sediment fused with lithium borate using isotope dilution laser ablation-inductively coupled plasma-mass spectrometry
Authors:Julien Malherbe  Fanny Claverie  Aitor Alvarez  Beatriz Fernandez  Rosario Pereiro  John L Molloy
Institution:1. Chemical Sciences Division, Material Measurement Laboratory, National Institute of Standards and Technology, 100 Bureau Drive, Stop 8391, Gaithersburg, MD 20899, USA;2. Department of Physical and Analytical Chemistry, Faculty of Chemistry, University of Oviedo, Julian Clavería, 8, 33006 Oviedo, Spain
Abstract:Quantitative analysis using laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) remains challenging primarily due to the lack of appropriate reference materials available for the wide variety of samples of interest and to elemental fractionation effects. Isotopic dilution mass spectrometry (IDMS) is becoming the methodology of choice to address these issues because the different isotopes of an element represent near-perfect internal standards. In this work, we investigated the lithium borate fusion of powdered solid samples, including soils, sediments, rock mine waste and a meteorite, as a strategy to homogenously distribute, i.e. equilibrate the elements and the added isotopically enriched standards. A comparison of this methodology using two pulsed laser ablation systems (ArF* excimer and Nd:YAG) with different wavelengths as well as two ICP-MS instruments (quadrupole and double-focusing sector field) was performed. Emphasis was put on using standard equipment to show the potential of the proposed strategy for its application in routine laboratories. Cr, Zn, Ba, Sr and Pb were successfully determined by LA-ICP-IDMS in six Standard Reference Materials (SRMs) representing different matrices of environmental interest. Experimental results showed the SRM fused glasses exhibited a low level of heterogeneity (intra- and inter-sample) for both natural abundance and isotopically enriched samples (RSD <3%, n = 3, 1σ). A good agreement between experimental results and the certified values was also observed.
Keywords:Laser ablation inductively coupled plasma mass spectrometry  Trace element  Isotope dilution mass spectrometry  Geological reference materials
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号