首页 | 本学科首页   官方微博 | 高级检索  
     


Multicommutated flow analysis system for determination of creatinine in physiological fluids by Jaffe method
Authors:Łukasz Tymecki  Jakub KorszunKamil Strzelak  Robert Koncki
Affiliation:University of Warsaw, Department of Chemistry, Pasteura 1, 02-093 Warsaw, Poland
Abstract:For the needs of photometric determination of creatinine according to Jaffe protocol a dedicated paired emitter detector diode (PEDD) detector has been developed. This PEDD device has been constructed in the compact form of flow-through cell (30 μL total volume and 7 mm optical pathlength) integrated with 505 nm LED-based emitter and 525 nm LED-based detector compatible with multicommutated flow analysis (MCFA) system. This fully mechanized MCFA system configured of microsolenoid valves and pumps is operating under microprocessor control. The developed analytical system offers determination of creatinine in the submillimolar range of concentrations with detection limit at ppm level. The throughput offered by the system operating according to multi-point fixed-time procedure for kinetic measurements is 15–40 samples per hour depending on the mode of measurements. The developed PEDD-based MCFA system has been successfully applied for the determination of creatinine in real samples of human urine as well as serum. The developed sampling unit used the system is free from effects caused by differences in sample viscosity.
Keywords:Flow analysis   Paired emitter detector diode   Creatinine   Jaffe method   Human urine and serum analysis   Clinical diagnostics
本文献已被 ScienceDirect 等数据库收录!