Spectroscopic set‐up for simultaneous UV‐Vis/(Q)EXAFS in situ and in operando studies of homogeneous reactions under laboratory conditions |
| |
Authors: | Matthias Bauer Gerhard Heusel Stefan Mangold Helmut Bertagnolli |
| |
Abstract: | A novel experimental set‐up for in operando studies of homogeneous catalyzed reactions under laboratory conditions has been developed and tested. It combines time‐resolved X‐ray absorption spectroscopy with UV/Vis spectroscopy. The reaction solution is stirred in a vessel and pumped in a circle by a peristaltic free gear‐wheel through a measurement cell. The X‐ray and UV/Vis beams probe the same sample volume of the cell orthogonally. Reactants can be added to the reaction mixture in the course of the measurements and a defined gas atmosphere can be adjusted up to a pressure of 10 bar. The in situ reduction of cerium(IV) ammonium nitrate to cerium(III) by isopropanol is studied as a test reaction with quick‐XANES and UV/Vis measurements with a time resolution of 60 s and 1 s, respectively. |
| |
Keywords: | two‐dimensional spectroscopy QEXAFS UV/Vis in operando |
|