Determination of trace impurities in chromium matrices after separation from Cr(III) using the oxalate form of anion exchanger
Authors:Venkateswarlu G  Sahayam A C  Chaurasia S C  Mukherjee T
Institution:National Centre for Compositional Characterisation of Materials (CCCM), B.A.R.C., E.C.I.L. (P.O.), Hyderabad 500062, India
Abstract:A method has been developed for the separation and determination of a set of 11 impurities from chromium matrices using oxalate form of Amberlite IRA 93. Due to slower kinetics of formation of the anionic complex, Cr(III) passed in the effluent while impurities forming strong complexes rapidly are retained on the exchanger. The adsorption of impurities of interest is found to be uniform in pH range 2-6. The adsorbed impurities are eluted with 2 mol l−1 HNO3 and determined by inductively coupled plasma-optical emission spectrometer (ICP-OES). The percentage recoveries of Al, Bi, Cd, Co, Cu, Fe, Mn, Ni, Pb, Ga and Zn are in the range 88-101% and separation of matrix is >99.9%. The method has been applied for the analysis of two samples namely CrCl3·6H2O and Cr. The R.S.D. of the method is 5-6% at >10 μg g−1 level and ∼15% at <1 μg g−1 level. The process blank values are in the range sub-μg g−1 and detection limits are in ng g−1 range.
Keywords:Oxalate  Cr(III)  Separation  Chromium matrix  Anion exchanger  Amberlite IRA 93  ICP-OES
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏