Development of a new electrochemical hydride generator with tungsten wire cathode for the determination of As and Sb by atomic fluorescence spectrometry
Authors:Zhang Wang-Bing  Gan Wu-Er  Lin Xiang-Qin
Institution:Department of Chemistry, University of Science and Technology of China, Hefei, Anhui 230026, PR China
Abstract:A novel electrochemical hydride generator has been developed for the determination of As and Sb. This newly devised hydride generator is constructed from a flowing electrolytic cell, in which the tungsten wire is selected as cathode. Compared with some cathode material usually used in electrochemical hydride generator, the tungsten cathode is of better interference tolerance, corrosion-resistant and longer working time. The characteristics of the cathode material, hydride generating efficiency and interferences of concomitant have been studied in detail. The detection limits (3σ) of As and Sb in sample solution were 0.10 μg L−1 and 0.15 μg L−1, the precisions for 11 replicate measurements of 20 μg L−1 As and Sb were 1.3% and 1.7%. The electrochemical hydride generator coupled with atomic fluorescence spectrometry has been applied to the determination of total As and Sb in tobacco samples.
Keywords:Electrochemical hydride generation  Tungsten  Atomic fluorescence spectrometry  Arsenic  Antimony  Tobacco samples