Use of CR-39 solid state nuclear track detector in neutron personnel monitoring |
| |
Authors: | M. P. Dhairyawan P. K. Marathe O. P. Massand |
| |
Affiliation: | Personnel Monitoring Section, Bhabha Atomic Research Center, Mumbai-400094, India |
| |
Abstract: | Neutron sources like 241Am–Be, 239Pu–Be, 252Cf and 14.6 MeV neutron generator are being used in oil exploration industries as well as in research institutions. While handling these neutron sources, personnel may be exposed to neutrons. Also personnel working in reactors, accelerators may receive dose from neutrons. These exposed individuals need to be monitored regularly for measurement of neutron doses. The individual neutron doses can be estimated by using Kodak NTA films and CR-39 Solid State Nuclear Track Detector with a polyethylene radiator to increase sensitivity in front in holder. Nearly 1450 personnel are being monitored regularly throughout the country on a quarterly basis. In India, the monitoring system adopted for individual neutron dose estimation having energy from 100 keV and above is described in this paper. Background counts of 0.20 mSv could be measured with CR-39 SSNTD foil system and it has been successfully introduced for Fast Neutron Personnel Monitoring for the country. |
| |
Keywords: | Neutron monitoring Detectors CR-39 SSNTD Dose estimation |
本文献已被 ScienceDirect 等数据库收录! |
|