首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Phase Systems and Post-Column Dithizone Reaction Detection for the Analysis of Organo-Mercurials by HPLC
Authors:C H Gast  J C Kraak
Institution:Laboratory for Analytical Chemistry , University of Amsterdam , Nieuwe Achtergracht 166, Amsterdam, The Netherlands
Abstract:Abstract

The suitability of HPLC (normal and reversed phase adsorption) with UV or post-column reaction detection for the analysis of organomercurials was investigated systematically. The separation of organomercurials is best carried out on a reversed phase system with a C-8 bonded phase material as the stationary phase and acetonitrile-aqueous sodium bromide mixtures as the mobile phase.

The precision and detection limit of the method and the efficiency of the extraction procedure were established. For the alkylmercury compounds the lowest limit of detection (80ppb) was obtained with the dithizone reaction detection and for the phenylmercury compounds with UV detection (60ppb). A chromatogram of a spiked fish (2ppmHg) and a river water sample (50ppbHg) is shown.
Keywords:Stable cobalt  fish  Zeeman  graphite furnace AAS  trace element
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号