首页 | 本学科首页   官方微博 | 高级检索  
     检索      


An electrochemical sensor for p-aminophenol based on the mesoporous silica modified carbon paste electrode
Authors:Dong Sun  Xiaokun Li  Huajie Zhang  Xiafeng Xie
Institution:1. School of Pharmacy, Wenzhou Medical College , Wenzhou 325035 , China sun_dong11@163.com;3. School of Pharmacy, Wenzhou Medical College , Wenzhou 325035 , China
Abstract:A mesoporous silica was synthesised and used to modify the surface of carbon paste electrode (CPE). The electrochemical behaviours of p-aminophenol were investigated. Compared to the unmodified CPE, the mesoporous silica-modified CPE obviously lowers the oxidation potential of p-aminophenol, and remarkably increases its oxidation peak current. The effects of pH value, amount of mesoporous silica, accumulation potential and time were examined. As a result, a sensitive, rapid and convenient electroanalytical method was developed for p-aminophenol. The linear range is from 0.025?mg?L?1 to 3?mg?L?1, and the limit of detection is 0.01?mg?L?1 after 2-min accumulation. Finally, the method was successfully used to determine p-aminophenol in water samples.
Keywords:aminophenol  detection  mesoporous silica  electrochemical sensor
设为首页 | 免责声明 | 关于勤云 | 加入收藏