首页 | 本学科首页   官方微博 | 高级检索  
     


Oxidability Determination in Waste Waters Using an Automatic Titrator Based on a Multicommutated Unsegmented Flow System
Authors:Cristina M.N.V. Almeida  Rui A.S. Lapa  José L.F.C. Lima
Affiliation:CEQUP/Departamento de Química-Física Faculdade de Farmácia da Universidade do Porto Rua Anibal Cunha , 164, 4050-047, Porto, Portugal
Abstract:Abstract

The automatic titrator based on a multicommutated unsegmented flow system was applied to redox titrations and used for oxidability determination in waters analysis. This automatic titrator allows the attainment of complete titration curves, being the determination of titrand concentration performed without requiring any prior calibration. After sample treatment (oxidation step), the oxidability determination in waste water samples was accomplished by the automatic flow titrator (titration step). Repeated determinations of standard solutions gave a 3.5% RSD (n=10, 0.010M) for repeatability and a 3.2% RSD (n=2, 0.057M) for reproducibility. Samples results (n=9) were in good agreement (t-test) with those obtained with a reference procedure.
Keywords:Multicommutated flow system  automatic flow titrator  oxidability  waste water  spectro-photometry
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号