Preliminary study on aerosol particle addition calibration method for on-line quantitative analysis of airborne radioactive particles with ICP-MS
Authors:Yong Yang Su  Zhi Ming Li  Hong Bo Dong  Guo Qing Zhou  Li Hua Zhai  Jiang Xu
Institution:1. Department of Engineering Physics, Tsinghua University , Beijing, China;2. Northwest Institute of Nuclear Technology , Xi'an, China;3. Northwest Institute of Nuclear Technology , Xi'an, China
Abstract:An ambient aerosol concentration enrichment system coupled with ICP-MS for real-time monitoring of airborne radioactive particles is now under development. ICP-MS is very sensitive to sample introduction conditions, so it is necessary to develop an easy-use calibration method for on-line quantitative analysis in field application. In this paper, a calibration method using standard solution instead of monodisperse particles was established and validated preliminarily. First of all, four parameters for the method were determined experimentally, including: uptake flow rate and nebulisation efficiency of the Microconcentric nebuliser, nebulisation/transport efficiency of Aridus Desolvating Sample Introduction System, and Relative Sensitivity Factor between 159?Tb and 174?Yb. Then, monodisperse terbium nitrate particles were generated by a commercial Vibrating Orifice Aerosol Generator. Continuous aerosols of ytterbium nitrate droplets were nebulised from standard solution. They were mixed together, desolvated through the membrane dryer and introduced into ICP-MS for on-line analysis of terbium nitrate particles. The air sampled from nuclear environment was also introduced into ICP-MS to investigate the effect of flow rate on instrument responses. Finally, atom numbers of 159?Tb in discrete terbium nitrate particles were determined using the calibration method and compared to the calculated value. Results show that when air flow rate increase from 10?mL?min?1 to 100?mL?min?1, the ratio of 159?Tb ion count to 174?Yb ion intensity keeps constant although instrument sensitivity decreases by a factor of 25. The relative standard deviation of 159?Tb atom number measured is better than 18%. The discrepancy with the calculated value could be attributed to the over-estimation of atom number in the particles generated by VOAG because there was some liquid leakage in the VOAG.
Keywords:aerosol particle addition calibration method  on-line quantitative analysis  airborne radioactive particles  ICP-MS
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号