Real-time quantitative analysis of H2, He,O2, and Ar by quadrupole ion trap mass spectrometry
Authors:Ottens Andrew K  Harrison W W  Griffin Timothy P  Helms William R
Affiliation:Department of Chemistry, University of Florida, Gainesville 32611-7200, USA.
Abstract:The use of a quadrupole ion trap mass spectrometer (QITMS) for quantitative analysis of hydrogen and helium as well as of other permanent gases is demonstrated. Like commercial instruments, the customized QITMS uses mass selective instability; however, this instrument operates at a greater trapping frequency and without a buffer gas. Thus, a useable mass range from 2 to over 50 daltons (Da) is achieved. The performance of the ion trap is evaluated using part-per-million (ppm) concentrations of hydrogen, helium, oxygen, and argon mixed into a nitrogen gas stream, as outlined by the National Aeronautics and Space Administration (NASA), which is interested in monitoring for cryogenic fuel leaks within the Space Shuttle during launch preparations. When quantitating the four analytes, relative accuracy and precision were better than the NASA-required minimum of 10% error and 5% deviation, respectively. Limits of detection were below the NASA requirement of 25-ppm hydrogen and 100-ppm helium; those for oxygen and argon were within the same order of magnitude as the requirements. These results were achieved at a fast data recording rate, and demonstrate the utility of the QITMS as a real-time quantitative monitoring device for permanent gas analysis.