首页 | 本学科首页   官方微博 | 高级检索  
     


Application of atmospheric pressure chemical ionisation mass spectrometry in the analysis of barbiturates by high-speed analytical countercurrent chromatography
Authors:Jones Jonathan J  Kidwell Huw  Games David E
Affiliation:Mass Spectrometry Research Unit, University of Wales Swansea, Grove Building, Singleton Park, Swansea SA2 8PP, Wales, UK. 147437@swan.ac.uk
Abstract:Four barbiturates (barbital, allobarbital, phenobarbital and butalbital) were analysed using high-speed analytical countercurrent chromatography (HSACCC) and high-performance liquid chromatography (HPLC) interfaced with mass spectrometry, using negative mode atmospheric pressure chemical ionisation (APCI). The polar biphasic solvent system of butyronitrile/acetonitrile/water (1:1:1) was used, in the upper-stationary, lower-mobile mode of operation, at a flow rate of 1 mL/min and a rotational speed of 1200 rpm, equating to an applied "g"-field of 177 g. The fractional stationary phase retention (S(F)) was 0.58. Representative mass spectral data are presented from the HPLC and the HSACCC analyses. Structural information was obtained using source-induced fragmentation at increased source block voltages. The effect of increasing g-field on chromatographic resolution is illustrated using the binary base system of butyronitrile/water (1:1), under electrospray ionisation.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号