Abstract:A new method for preparation of sorbent-based ultra-thick film traps for concentration of trace volatile components from gaseous matrices is described. The procedure is based on blowing a prepolymer (polydimethylsiloxane) through a capillary tube, forming an irregular film of stationary phase. Subsequently, the prepolymer is immobilized in a few seconds by heating to 200 degrees C. Evaluation of the performance of the new traps showed that the loss of efficiency, compared to regular smooth film traps is only on the order of 20-30%. In terms of breakthrough volume, this loss in performance is rather insignificant. The technology is extremely simple and allows a rapid and cheap production of a large number of ultra-thick film traps, even in non-specialized laboratories. The method can be applied to any type of cross-linkable stationary phase, thereby expanding the scope of sorbent-based trapping and preconcentration concept. Many applications are anticipated in trace and ultra-trace analysis in a wide range of fields, such as environmental chemistry, polymers, food and process analysis.
Keywords:Open tubular traps  Sample handling  Sorbent trapping  Instrumentation  Volatile organic compounds