首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Direct-scanning alpha-spectrometer for americium and plutonium contamination on highly-enriched uranium surfaces
Authors:W C Ward  H E Martinez  C L Abeyta  A N Morgan  T O Nelson
Institution:(1) Energy and Process Engineering, MS-J576 Los Alamos National Laboratory, 87545 Los Alamos, NM, USA;(2) Nuclear Materials Technology, MS-E510 Los Alamos National Laboratory, 87545 Los Alamos, NM, USA
Abstract:Trace239Pu and241Am contamination on a surface whose alpha count is dominated by235U and234U decay has been successfully quantified by counting swipes in external alpha-spectroscopy chambers. The swipe process, however, is labor intensive and subject to uncertainties in the swiping process as well as degraded spectral resolution due to the presence of the swipe material. A multichannel instrument for automated in situ measurements of interior and exterior contamination has been developed which incorporates a rotary table, 13 fixed ion-implanted silicon detectors, and spectroscopy electronics. Custom software was written to allow alpha-spectrometer to function as a virtual instrument in the LabView environment. This system gives improved speed and resolution as well as a complete log of the location of areas of high surface contamination, a feature not practical to obtain by other methods, and one which opens the possibility of long term studies such as Pu outgrowth evaluation employing the instrument. We present performance data as well as system integration, calibration, control, and dynamic geometric efficiency calculations related to the design of this and next generation systems.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号