首页 | 本学科首页   官方微博 | 高级检索  
     


Performance of a new automatic carbon dioxide coulometer
Authors:Edward W. D. Huffman   jr.
Affiliation:Huffman Laboratories, Inc., 3830 High Court, P. O. Box 777, Wheat Ridge, Colorado 80033 USA
Abstract:The Coulometrics Inc. CO2 coulometer has been shown to be an accurate and reliable CO2 measuring device. The coulometric efficiency is essentially 100%. This means that the method can be considered as a standard reference method for CO2. As with a CO2 absorption tube, certain potential interferences must be considered, however, the removal of these interferences is well documented in the literature.The CO2 coulometer has found a variety of applications in the author's
3. effect of flow rate on absorption of CO2 by coulometer
CaCO3 (mg)C (mg)Flow rate (cm3/min)C found (mg)C (%)
27.9893.3591003.366912.03
28.6043.4322003.434312.00
29.2593.5113003.514912.01
33.8084.0574004.038111.94
5.6290.6755000.676012.01
10.3111.2375001.233711.96
15.6471.8785001.870611.95
35.2144.2265004.198211.92
40.7334.8885004.821211.84
59.6787.1615007.026311.77
30.3863.6467803.594111.83
29.7813.5747803.536111.87
28.1133.37411503.253411.57
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏