首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Effective on-line coupling of HPLC and flame-AAS by means of hydraulic high pressure nebulization
Authors:G Weber  H Berndt
Institution:(1) Institut für Spektrochemie und angewandte Spektroskopie, Bunsen-Kirchhoff-Str. 11, D-4600 Dortmund 1, FRG
Abstract:Summary Hydraulic high pressure nebulization is used as an effective way of on-line coupling of HPLC to flame-AAS for speciation of metal compounds in the ng range. Compared to coupling with a conventional nebulizer a signal enhancement by a factor of 7.5 (peak height) and 10.1 (peak area) for copper is obtained. Using an injected volume of 50 μL the detection limits for Cu, Fe, Ni and Cd are below 0.1 μg/mL (=5ng) and for Mg below 0.01 μg/mL (=0.5ng). The effects of HPLC flow-rate and nebulization nozzle diameter on the signal peak height have been investigated. The performance of the system is demonstrated using speciation of iron. A base-line separation of Fe(II) and Fe(III) is achieved within two minutes. Also species changes (Fe acetate to Fe citrate) can be analysed using the proposed system.
Keywords:Column liquid chromatography  On-line flame-AAS detection  Hydraulic high pressure nebulization  Trace metal speciation
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏