Determination of Mercury in Urine and Seawater by Flow Injection Vapor Generation Isotope Dilution Inductively Coupled Plasma Mass Spectrometry
Authors:Ming-Tsai Wei  Shiuh-Jen Jiang
Abstract:A simple and inexpensive laboratory-built vapor generator was used with inductively coupled plasma mass spectrometry (ICP-MS) for the determination of mercury in urine and seawater samples. The applications of vapor generation ICP-MS alleviated the non-spectroscopic interferences and the sensitivity problem of mercury determination encountered when the conventional pneumatic nebulizer was used for sample introduction. The concentration of mercury was determined by isotope dilution method. The isotope ratio of mercury was calculated from the peak areas of each injection peak. The repeatability of the peak areas and isotope ratio determinations of seven consecutive injections of 1 ng mL?1 Hg solution were 2.3% and 2.2%, respectively. This method has a detection limit of 0.07 ng mL?1 for mercury. This method was applied to determine mercury in a CASS-3 nearshore seawater reference sample, NASS-4 open ocean seawater reference sample, NIST SRM 2670 freeze-dried urine reference sample and several urine and seawater samples collected from National Sun Yat-Sen University. The results for the reference samples agreed satisfactorily with the reference values. Results for other samples analyzed by the isotope dilution method and the method of standard additions agreed satisfactorily. Precision was better than 10% for most of the determinations.
Keywords:Inductively coupled plasma mass spectrometer  Mercury  Flow injection  Vapor generation  Isotope dilution  Urine and seawater sample
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号