Dosimetric properties of the newly developed KLT-300 (LiF:Mg,Cu,Na,Si) TL detector
Authors:Lee J I  Kim J L  Chang S Y  Chung K S  Choe H S
Institution:

a Health Physics Department, Korea Atomic Energy Research Institute, P.O. Box 105, Yuseong, Daejeon 305-600, South Korea

b Department of Physics and Physics Education, Gyeong-Sang National University, Chinju 660-701, South Korea

Abstract:The dosimetric properties of the newly developed KLT-300 (KAERI LiF:Mg,Cu,Na,Si TL detector) in KAERI (Korea Atomic Energy Research Institute) were investigated. The sensitivity of the TL detector was about 30 times higher than that of the TLD-100 by light integration. In the study of the dose linearity of the detector, the dose response was very linear up to 10 Gy and a sublinear response was observed at higher doses. The energy response of the detector was studied for photon energies from 20 to 662 keV. The results show that a maximum response of 1.004 at 53 keV and a minimum response of 0.825 at 20 keV were observed. The reproducibility study for the TL detector was also carried out. The coefficients of variation for each detector separately did not exceed 0.016, and for all the 10 detectors collectively it was 0.0054. IEC Standard requires that the coefficient of variation shall not exceed 0.075. So, the reproducibility of this new TL detector sufficiently satisfied the IEC requirements. A detection threshold of the detector was investigated and found to be 70 nGy by Harshaw 4500 TLD Reader.
Keywords:LiF:Mg  Cu  Na  Si  Thermoluminescence  TLD  KLT-300